The SafeHood is a recirculating fume cupboard designed to handle chemical vapours & aerosols, and fine airborne powders in the laboratory environment.
- Manufactured in accordance with BS 7989, Afnor NFX 15211 and Etaf type 1 and 2.
- Maximes energy saving on laboratory air conditioning in comparison to ducted models.
- Provides an economic solution for low volume applications when a ducted extract system is not viable and necessary.
- No influence on the balancing and control of air pressure levels in the laboratory.
- Extra-large activated carbon filters with optimised granularity for the highest chemical specificity.
- Microprocessor controlled airflow with average inflow speed display, airflow alarm and filter efficiency.
- Unique filter clamping mechanism provides the highest air tightness between the filters and sealing face, together with a simple filter removal procedure.
- Large choice of specific application activated carbon filters.
- Choice of polypropylene, ceramic or stainless steel work surface with liquid containment.
Example applications:
- Handling chemical substances in the laboratory.
- Sample preparation for anathomo-pathology.
- Rigid or flexible endoscope decontamination.
- Handling of adhesives and solvents.
- Handling of airborne powders and chemical aerosols.
Comfortable
- Smooth operating sliding sash glass window provides easy access to the work area.
- The glass side walls offer excellent visibility and ease of cleaning.
- Integral fluorescent lighting.
- Extremely low noise level induction exhaust fan.
Safe
- Highest containment factor provided by the most sophisticated microprocessor controlled air speed regulation that maintains the barrier average velocity at 0.5 m/sec, independent of the sliding sash position.
- Visual and acoustic alarm for low inflow velocity and sash position.
- All electrical components are isolated from the air flow.
- Optional dual exhaust safety filter (4kg).
- Safety slot(s) on the front panel for visual filter(s) identification (in according with European safety regulation).
- Pre-filters are provided as standard and are easily removable from inside the cabinet providing the highest operator safety.
Efficient
- Extra-large activated carbon filters with optimised granularity for the highest chemical specificity and efficiency.
- Large choice of specific application filters.
- Front working opening aerodynamically designed to ensure minimum turbulence whilst providing maximum containment.
- Reduced speed function for standby application.
Technical Specification:
For technical information on the range of available carbon filters please click here:
{Charcoal filters for SafeHood}
SafeHood 75 Order Code FS200000 |
SafeHood 120 Order Code FS100000 |
SafeHood 165 Order Code FS300000 |
|
Dimensions: (W x D x H) mm |
Work Chamber: 670 x 550 x 600 Overall Size: 750 x 720 x 1200 Height on Stand: 2050 |
Work Chamber: 1120 x 550 x 600 Overall Size: 1200 x 720 x 1200 Height on Stand: 2050 |
Work Chamber: 1570 x 550 x 600 Overall Size: 1650 x 720 x 1200 Height on Stand: 2050 |
Construction: |
Carcase: Epoxy painted cold rolled steel Viewing Panels: Toughen Safety Glass Worksurface: AISI 304 Stainless Steel, Ceramic or Polypropylene |
||
Volume Flow: (m3/hr) |
300 |
550 |
800 |
Face Velocity: (m/s) |
> 0.4 m/s (set point 0.5 m/s) |
||
Main Carbon Filter: |
1 x 12 kg |
2 x 12 kg |
3 x 12 kg |
Safety Filter: (optional) |
1 x 4 kg |
2 x 4 kg |
3 x 4 kg |
Weight: (excluding filters) |
95 kg |
130 kg |
180 kgs |
Power Supply: |
220 / 240 V – 50 Hz |
||
Power: |
420 Watts |
800 Watts |
1170 Watts |
Noise Level: |
< 57 dB (A) |
< 59 dB (A) |
|
Lighting Level: |
>800 lux |
>800 lux |
>800 lux |
Options & Accessories:
Safehood 75 |
Safehood 120 |
Safehood 165 |
|
Support Stand |
AS44000 |
AS43000 |
AS45000 |
Polypropylene work surface |
AZ90001 |
AZ90101 |
AZ90201 |
Stainless steel work surface |
AZ90000 |
AZ90100 |
AZ90200 |
Ceramic work surface |
AZ90003 |
AZ90103 |
AZ90203 |
Additional vacuum tap |
AZZA000 |
||
Additional gas tap |
AZZ0001 |